Raditeq PRB106FZ ISO17025 accredited (RvA) calibration certificate for RadiSense 26/40 GHz

Specifications

ISO17025 accredited calibration for RadiSense® 2010 series

ISO17025 accredited calibration performed at Kiwa Dare Calibrations in accordance to the international standard IEC 61000-4-3 (2006) 3e ed. + A1 (2007) + A2 (2010)

The polarisation of the E-field is vertical.

The E-field is measured by each separate axis.

 Measured frequency points:

  • Min/Max level: > 3 V/m and < 200 V/m till 150 MHz
  • Min/Max level: > 3 V/m and < 100 V/m from 150 MHz
  • 10 kHz – 100 kHz: 10 kHz steps
  • 100 kHz – 1 MHz: 100 kHz steps
  • 1 MHz – 10 MHz: 1 MHz steps
  • 10 MHz – 100 MHz: 10 MHz steps
  • 100 MHz – 1 GHz: 50 MHz steps
  • 1 GHz – 6 GHz: 200 MHz steps
  • 6 GHz – 12.5 GHz: 500 MHz steps

Calibrated under RvA accreditation:

  • Frequency response three probe orientations @ 20V/m
  • Linearity for three probe orientations @ 100 MHz from – 6dB to + 6dB of the reference field level (1 dB steps)
  • Anisotropy three probe orientations @ 100kHz; 1MHz; 10 MHz; 100MHz; 1GHz and 10GHz.

The calibration certificate is provided in PDF format and all calibration correction data is delivered in Excel format.

Product Number:

RQ-PRB106FZ

Manufacturer:

Raditeq

Quantity:

Total
Unit price
Base Product
Availability:
Call or email for delivery time
Description
Additional Informations
Reviews
Query on the Item
Do you have any questions on this item?
Your E-mail Address
My Question:
Please retype the letters shown here
captcha
Description
Reviews
Query on the Item
Do you have any questions on this item?
Your E-mail Address
My Question:
Please retype the letters shown here
captcha

These items were last seen: